												It is possible to calibrate DISK probe using the Calibration Wizard.
The implementation at version 3.6 requires a manual modification of the program in order to have the disk calibrated.
The procedure to be followed is:
| Create a calibration Wizard of the disk probe selecting the graphical definition and all the required angles to be calibrated. | 
 | 
| The Wizard creates a standard definition of the probe, without any definition of the disk, that shall be add manually | 
 | 
In order define the disk probe it is necessary to add, as last parameter on each disk definition, the following code ,DISK,height where height represent the height of the disk | 
 | 
| Consider, for a good calibration to let the sphere to be measure with at least 9 points | 
 |